IRMPD in 3D Quadrupole Ion Trap -- [E09]
Infrared spectroscopy of ions integrated with mass spectrometry/ESQUIRE 3000+
The room is equipped with a Paul trap mass spectrometer (Esquire Bruker 3000+) that has been extensively modified to couple with the CLIO free electron laser, an OPO/OPA tabletop laser, a CO2 laser, and an UV laser. An ion mobility device (DIMS) is also available on the ion trap. This setup allows for IR or UV spectra of ions to be derived, with isomeric forms separated by differential ion mobility beforehand. Ions are gas-phase ionized via an ESI source, and the ion trap is modified to introduce neutral molecules for ion-molecule reactions.
Disciplines
-
-
-
-
-
-
-
- Molecular Physicochemistry
- Analytical Chemistry
-
-
-
-
-
-
Complementary Techniques
-
-
-
-
-
-
-
- Tandem mass spectrometry (MS2 or MSn)
- UVPD spectroscopy (UV photodissociation)
- Ion mobility of the DIMS type
-
-
-
-
-
-
Applications
-
-
-
-
-
-
-
- Characterization of reaction mechanisms via spectroscopy of reaction intermediates.
- Characterization of products in a complex mixture.
- Characterization of the nature of oxidation and fragmentation products of peptides.
- Characterization of chiral discrimination phenomena in complexes through UV-induced photofragmentation.
- Characterization of isomeric species within a mixture.
- Study of ion-molecule reactions of interstellar interest.
-
-
-
-
-
-
Sample Environment
-
-
-
-
-
-
-
- Electrospray source
- Solution concentration ~ 10-5 to 10-6 M, typically in a water:methanol mixture
- Possibility of introducing solid, liquid, or neutral gas samples into the ion trap.
-
-
-
-
-
-
Energy Range
Infrared :
-
-
-
-
-
-
-
- 5 to 20 µm (CLIO)
- 2.5 to 5 µm (OPO/OPA)
-
-
-
-
-
-
UV : 266 nm (soon to be 213 and 355 nm)
Detectors / Complementary Equipment
-
-
-
-
-
-
-
- Infrared-sensitive detectors:
- OPO/OPA tabletop laser (2.5-5 µm)
- UV laser (Litron) (@ 266 nm)
- CO2 laser (10.6 µm)
- Differential Ion Mobility Spectrometry (DIMS)
-
-
-
-
-
-
Beam Size
Adjustable.
The infrared beam is focused within the annular electrode, which has been pierced with a hole approximately 2 mm in diameter.
Technical Details
-
-
-
-
-
-
-
- The mass spectrometer is of the Paul trap type (Bruker Esquire 3000+) equipped with an electrospray source. It is a quadrupole trap (Paul trap) coupled with an atmospheric pressure ESI source. Resolution is approximately 5000 at m/z 3000. Pressure is around 10^-5 Torr.
- OPO/OPA laser (Laser Vision). The system consists of a Nd pump laser, followed by an OPO and an OPA. KTP (KTiOPO4) and KTA (KTioAsO4) crystals are used to successively change the starting wavelength until the desired one is achieved.
- UV Laser TRLi 850 (Litron). The laser, with an output energy at 1064 nm (approximately 130 mJ/pulse), can be coupled with second, third, fourth, and fifth harmonic modules.
-
-
-
-
-
-
Contact
Head of Experiments : Debora Scuderi